Temperature and Humidity

Temperature Calibration

Calibration certificate
A moment from calibration work

What Is Calibration?

Temperature calibration determines the difference between a temperature instrument's reading and a reference with documented traceability. Measurement is done either by comparison in a controlled environment such as a dry block, liquid bath or climate chamber, or by using fixed points.

Why Is It Done?

The aim is for temperature-dependent decisions to rest on a reliable measurement. In heat treatment, sterilisation, cold chain and food safety processes, temperature directly determines product conformity; a thermometer with unknown deviation puts all of these processes in doubt.

Which Instruments?

  • Digital and analogue thermometers, immersion probes
  • Thermocouples and resistance thermometers (PT100, PT1000)
  • Infrared (non-contact) thermometers
  • Ovens, furnaces, incubators and autoclaves
  • Climate and stability chambers
  • Cold stores and refrigerators
  • Data loggers

Measured Quantities

  • Indication error: the difference between the instrument's reading and the reference
  • Repeatability: the spread of repeated readings at the same point
  • Temperature distribution within the volume (in chambers and ovens): the difference between the hottest and coldest points
  • Temporal stability: fluctuation around the set point
  • The effect of emissivity on infrared instruments

Calibration Interval

The calibration interval is not fixed by any standard and there is no general rule such as "once a year". The interval is set by weighing together the manufacturer's recommendation, how often and under what conditions the instrument is used, how critical the measurement is, the drift shown by the instrument's previous calibration results, and any legal requirement. ILAC-G24 / OIML D10 describe the methods for setting it. The sound approach is to start with a reasonable interval for the first period and then lengthen or shorten it in the light of subsequent results: an instrument that proves stable can have its interval extended, one that drifts has it shortened.

Method and Traceability

  • ISO/IEC 17025
  • EURAMET cg-11 (digital thermometers)
  • EURAMET cg-13 (thermocouple calibration)
  • IEC 60751 (resistance thermometers)
  • ITS-90 international temperature scale
  • JCGM 100 (GUM — uncertainty)

Method and Traceability

  • ISO/IEC 17025
  • EURAMET cg-11 (digital thermometers)
  • EURAMET cg-13 (thermocouple calibration)
  • IEC 60751 (resistance thermometers)
  • ITS-90 international temperature scale
  • JCGM 100 (GUM — uncertainty)

Calibration Process

  1. 1Receipt of the Instrument and Initial InspectionThe instrument is received and its physical condition and functioning are checked. Any damage or fault is reported before calibration; calibrating a faulty instrument yields no meaningful result.
  2. 2Conditioning to Ambient ConditionsThe instrument is left until it reaches equilibrium with the laboratory environment. A measurement taken before the temperature difference from transport has settled is erroneous.
  3. 3Preparation of the ReferenceA reference thermometer, dry block or climatic chamber with documented traceability is brought to a stable condition.
  4. 4Selection of the Measurement PointsPoints representing the instrument's working range are selected; where the customer has stated a critical operating point, that point is always included.
  5. 5Reading at Steady StateAt each point, thermal equilibrium between reference and instrument is awaited and repeated readings are taken. This waiting time is the most critical step in thermal calibration.
  6. 6Calculation of the DeviationThe difference between the instrument's indication and the reference value is calculated for each point.
  7. 7Building the Uncertainty BudgetThe expanded uncertainty is calculated by combining the contributions of the reference uncertainty, repeatability, resolution, homogeneity and stability.
  8. 8Issuing the CertificateMeasurement results, deviations and uncertainty values are entered on the certificate; where requested, a statement of conformity is added using an agreed decision rule.

Measurement Uncertainty

Every calibration result is given together with a measurement uncertainty; a result reported without its uncertainty is incomplete. The uncertainty is budgeted following the GUM (JCGM 100) approach: the uncertainty of the reference, repeatability, resolution, environmental effects and the stability of the instrument are combined. The certificate states the expanded uncertainty with the coverage factor (usually k = 2) and the approximate level of confidence. The smallest uncertainty a laboratory can report is limited by its calibration and measurement capability (CMC) in the accredited scope; no better uncertainty may be declared.

  • The reference thermometer's calibration uncertainty
  • The reference's stability (drift) since its calibration
  • Temperature homogeneity in the dry block or bath
  • Temporal stability and fluctuation
  • The instrument's resolution
  • Repeatability
  • The heat-conduction effect caused by immersion depth

Certificate Content

  • Laboratory details and certificate number
  • The customer and identification of the instrument calibrated: make, model, serial number
  • Date of calibration and date of issue of the certificate
  • The calibration method applied or reference to the procedure
  • Reference standards used and traceability information
  • Ambient conditions during calibration (temperature, humidity and, where relevant, pressure)
  • Measurement points, reference values and instrument readings
  • The deviations calculated
  • Expanded measurement uncertainty and coverage factor
  • The personnel who performed and who approved the calibration
  • The method used: comparison or fixed point
  • Immersion depth and probe placement
  • A location diagram of the measurement points for chamber calibrations

Conformity Statement and Limits

A calibration certificate is NOT a certificate of conformity. Calibration measures and reports the deviation of the instrument; it does not correct that deviation and does not say that the instrument is "correct". To decide whether an instrument is fit for use, a tolerance (acceptance criterion) must be defined — and it is the user who sets that tolerance, not the laboratory. If a statement of conformity is requested, ISO/IEC 17025 clause 7.8.6 requires a DECISION RULE to be agreed in advance, defining how measurement uncertainty is taken into account in the decision; a conformity statement made without regard to uncertainty is misleading. The certificate also shows the condition AT THE TIME of calibration: later use, transport or shock can change the instrument's condition, and the certificate does not cover that. Where a next calibration date appears on the certificate, it is not the laboratory's decision but a reminder entered according to the interval stated by the customer.

Accreditation

Calibration is carried out under ISO/IEC 17025, which requires the laboratory's technical competence, impartiality and metrological traceability. Accreditation is granted per METHOD and per MEASUREMENT RANGE: a quantity or range that does not appear in the laboratory's scope annex is not accredited, and no certificate bearing the accreditation mark may be issued for that measurement. The scope also states the smallest uncertainty the laboratory can offer (CMC); no better uncertainty may be declared. EUROLAB holds internationally recognised accreditation under ISO/IEC 17025, granted by accreditation bodies that are signatories to the ILAC mutual recognition arrangement; whether your instrument can be calibrated within accredited scope is confirmed in writing at the quotation stage.

Scope and authorisations

Related Calibrations

Frequently Asked Questions

Is a single point enough for oven or chamber calibration?

No. Temperature is not distributed homogeneously within a volume; the area near the door, the shelf edges and the fan zone can differ markedly. Multiple points distributed through the volume are therefore measured, and the hottest/coldest points are reported. In heat-treatment applications, this is separately required as a temperature uniformity survey (TUS).

Why does my infrared thermometer read differently?

An infrared instrument does not measure temperature directly; it measures the radiation from a surface and converts it to temperature using an emissivity value. Emissivity is low on bright metal surfaces, and the instrument reflects the surrounding heat. Calibration is done at a specific emissivity value; if this value is not set correctly in use, the deviation on the certificate does not apply.

Why does immersion depth matter?

If a probe is not immersed sufficiently, it conducts heat outward along its stem and the temperature at its tip deviates from the true ambient temperature. This effect grows especially large at high temperatures. The immersion depth used in calibration is stated on the certificate, and the same depth is expected to be observed in use.

Does calibration correct my instrument?

No. Calibration MEASURES and reports the deviation. Correcting the deviation is a separate operation (adjustment) and must be requested. When an adjustment is made, correct practice is to report the results before and after adjustment separately.

Should the certificate state the next calibration date?

Determining the calibration interval is the user's responsibility; the laboratory cannot decide this interval. If a date appears on the certificate, it is a reminder set according to the interval the customer stated, not a technical assertion by the laboratory.