What Is Calibration?
Force calibration compares a force-measuring instrument's reading against a reference force with documented traceability. For materials testing machines this is done under ISO 7500-1, and the machine is assigned an accuracy class.
Why Is It Done?
The aim is for strength values obtained from materials tests to be reliable. An error in a tensile-test result directly affects the material-acceptance decision.
Which Instruments?
- Tension-compression (universal) testing machines
- Load cells
- Force gauges and dynamometers
- Hardness testers (the force-application system)
- Presses and compression testers
- Proving rings and reference transducers
Measured Quantities
- Indication error: with increasing and decreasing force
- Repeatability and reproducibility
- Zero-return error
- Hysteresis (return error)
- Resolution
- Assigning the accuracy class (ISO 7500-1: Class 0.5 / 1 / 2, etc.)
Calibration Interval
The calibration interval is not fixed by any standard and there is no general rule such as "once a year". The interval is set by weighing together the manufacturer's recommendation, how often and under what conditions the instrument is used, how critical the measurement is, the drift shown by the instrument's previous calibration results, and any legal requirement. ILAC-G24 / OIML D10 describe the methods for setting it. The sound approach is to start with a reasonable interval for the first period and then lengthen or shorten it in the light of subsequent results: an instrument that proves stable can have its interval extended, one that drifts has it shortened.
Method and Traceability
- ISO/IEC 17025
- ISO 7500-1 (materials testing machines)
- ISO 376 (force-measuring transducers)
- EN 12390-4 (concrete compression testing machines)
- JCGM 100 (GUM — uncertainty)
Method and Traceability
- ISO/IEC 17025
- ISO 7500-1 (materials testing machines)
- ISO 376 (force-measuring transducers)
- EN 12390-4 (concrete compression testing machines)
- JCGM 100 (GUM — uncertainty)
Calibration Process
- 1Receipt of the Instrument and Initial InspectionThe physical condition of the instrument is examined for wear and damage. On dimensional instruments the cleanliness of the measuring faces is also checked.
- 2Conditioning to Ambient ConditionsThe instrument is brought to equilibrium at laboratory temperature. In dimensional measurement a temperature difference turns directly into measurement error through thermal expansion of the material.
- 3Preparation of the Reference StandardsWeights, gauge blocks, a dead-weight tester or a reference transducer with documented traceability are prepared.
- 4Preloading and Zero CheckIn force, torque and weighing calibrations the instrument is taken up to full load and back down several times; the zero point is then checked.
- 5Readings at the Measurement PointsReadings are taken in increasing and decreasing directions at points representing the measuring range; hysteresis is determined in this way.
- 6Repeatability MeasurementsThe same point is measured repeatedly; in weighing calibration an eccentricity (off-centre load) check is also performed.
- 7Application of CorrectionsCorrections for local gravity, air buoyancy and temperature are taken into account; ignoring them produces significant deviation in precision measurements.
- 8Uncertainty and CertificateThe uncertainty budget is established and the results are entered on the certificate; where requested, a statement of conformity is added with a decision rule.
Measurement Uncertainty
Every calibration result is given together with a measurement uncertainty; a result reported without its uncertainty is incomplete. The uncertainty is budgeted following the GUM (JCGM 100) approach: the uncertainty of the reference, repeatability, resolution, environmental effects and the stability of the instrument are combined. The certificate states the expanded uncertainty with the coverage factor (usually k = 2) and the approximate level of confidence. The smallest uncertainty a laboratory can report is limited by its calibration and measurement capability (CMC) in the accredited scope; no better uncertainty may be declared.
- The reference force transducer's uncertainty and drift
- The temperature effect and the transducer's temperature coefficient
- The effect of off-axis (side) loading
- Repeatability and hysteresis
- Instrument resolution
- The stiffness of the connecting fixtures
Certificate Content
- Laboratory details and certificate number
- The customer and identification of the instrument calibrated: make, model, serial number
- Date of calibration and date of issue of the certificate
- The calibration method applied or reference to the procedure
- Reference standards used and traceability information
- Ambient conditions during calibration (temperature, humidity and, where relevant, pressure)
- Measurement points, reference values and instrument readings
- The deviations calculated
- Expanded measurement uncertainty and coverage factor
- The personnel who performed and who approved the calibration
- The calibrated force range and direction (tension/compression)
- The assigned accuracy class (if ISO 7500-1 was applied)
- The connecting fixtures used
- Which column/crosshead position of the machine the measurement was made at
Conformity Statement and Limits
A calibration certificate is NOT a certificate of conformity. Calibration measures and reports the deviation of the instrument; it does not correct that deviation and does not say that the instrument is "correct". To decide whether an instrument is fit for use, a tolerance (acceptance criterion) must be defined — and it is the user who sets that tolerance, not the laboratory. If a statement of conformity is requested, ISO/IEC 17025 clause 7.8.6 requires a DECISION RULE to be agreed in advance, defining how measurement uncertainty is taken into account in the decision; a conformity statement made without regard to uncertainty is misleading. The certificate also shows the condition AT THE TIME of calibration: later use, transport or shock can change the instrument's condition, and the certificate does not cover that. Where a next calibration date appears on the certificate, it is not the laboratory's decision but a reminder entered according to the interval stated by the customer.
Accreditation
Calibration is carried out under ISO/IEC 17025, which requires the laboratory's technical competence, impartiality and metrological traceability. Accreditation is granted per METHOD and per MEASUREMENT RANGE: a quantity or range that does not appear in the laboratory's scope annex is not accredited, and no certificate bearing the accreditation mark may be issued for that measurement. The scope also states the smallest uncertainty the laboratory can offer (CMC); no better uncertainty may be declared. EUROLAB holds internationally recognised accreditation under ISO/IEC 17025, granted by accreditation bodies that are signatories to the ILAC mutual recognition arrangement; whether your instrument can be calibrated within accredited scope is confirmed in writing at the quotation stage.
Scope and authorisationsRelated Calibrations
Frequently Asked Questions
Should the tension and compression directions be calibrated separately?
Yes, if the machine is used in both directions. When the loading direction changes, the behaviour of the fixtures and mechanical clearances differ; calibration done in one direction is not considered valid for the other.
What does the accuracy class mean?
ISO 7500-1 assigns the machine to a class based on its measured deviation, repeatability and zero error. The test standard to be applied usually requires a minimum class; for example, in metal tensile tests, a result obtained with a machine below a certain class is not accepted.
Does calibration correct my instrument?
No. Calibration MEASURES and reports the deviation; correcting it is a separate operation (adjustment) and must be requested separately. When an adjustment is made, correct practice is to report the before- and after-adjustment results separately.
Can I tell from the certificate whether my instrument exceeds tolerance?
The certificate gives the deviation and the uncertainty; you determine the tolerance. If you want a statement of conformity, you must state your acceptance criterion and decision rule beforehand; a pass/fail decision given without accounting for uncertainty is misleading.