Acoustic Calibration

Calibration certificate
A moment from calibration work

What Is Calibration?

Acoustic calibration compares the sound-measurement chain (microphone, preamplifier and measuring instrument) against a reference with documented traceability. For sound level meters, this is carried out as periodic verification under IEC 61672-3, and whether the instrument's class is retained is assessed.

Why Is It Done?

The aim is for noise measurements to be legally and technically valid. Workplace noise exposure, environmental noise and product noise declarations rest directly on these instruments and are often subject to enforcement.

Which Instruments?

  • Sound level meters (Class 1 and Class 2)
  • Acoustic calibrators and pistonphones
  • Noise dosimeters
  • Measurement microphones and preamplifiers
  • Vibration meters (together with the acoustic chain)
  • Sound-power measurement systems

Measured Quantities

  • Acoustic sensitivity and level accuracy
  • Frequency-weighting response (A, C, Z)
  • Time-weighting response (Fast, Slow, Impulse)
  • Linearity: across the measuring range
  • Its own noise floor
  • Directivity (microphone)
  • Whether the instrument class (Class 1 / Class 2) is retained

Calibration Interval

The calibration interval is not fixed by any standard and there is no general rule such as "once a year". The interval is set by weighing together the manufacturer's recommendation, how often and under what conditions the instrument is used, how critical the measurement is, the drift shown by the instrument's previous calibration results, and any legal requirement. ILAC-G24 / OIML D10 describe the methods for setting it. The sound approach is to start with a reasonable interval for the first period and then lengthen or shorten it in the light of subsequent results: an instrument that proves stable can have its interval extended, one that drifts has it shortened.

Method and Traceability

  • ISO/IEC 17025
  • IEC 61672-1 (sound level meter requirements)
  • IEC 61672-3 (periodic verification)
  • IEC 60942 (acoustic calibrators)
  • IEC 61252 (dosimeters)
  • JCGM 100 (GUM — uncertainty)

Method and Traceability

  • ISO/IEC 17025
  • IEC 61672-1 (sound level meter requirements)
  • IEC 61672-3 (periodic verification)
  • IEC 60942 (acoustic calibrators)
  • IEC 61252 (dosimeters)
  • JCGM 100 (GUM — uncertainty)

Calibration Process

  1. 1Receipt of the Instrument and Initial InspectionThe cleanliness and physical condition of the instrument are checked. On volumetric glassware, cracks and wear are examined; on flow meters, residues.
  2. 2Conditioning of Environment and FluidThe instrument and the fluid to be used are brought to equilibrium under laboratory conditions. Temperature directly affects the density of the fluid and therefore the result.
  3. 3Preparation of the Reference SystemA reference flow meter, gravimetric rig or acoustic calibrator with documented traceability is put into service.
  4. 4Selection of the Measurement PointsPoints representing the instrument's operating range are selected; in flow calibration the low-flow region is addressed separately because error grows there.
  5. 5Taking the MeasurementsAt each point, repeated measurements are taken once a steady flow or steady signal has been established.
  6. 6Application of CorrectionsTemperature, pressure and density corrections are applied; in the gravimetric method the air buoyancy correction is taken into account as well.
  7. 7Uncertainty BudgetThe reference uncertainty, repeatability, temperature stability and reading resolution are combined.
  8. 8Issuing the CertificateThe results and uncertainty values are entered on the certificate; a statement of conformity with the decision rule is added if requested.

Measurement Uncertainty

Every calibration result is given together with a measurement uncertainty; a result reported without its uncertainty is incomplete. The uncertainty is budgeted following the GUM (JCGM 100) approach: the uncertainty of the reference, repeatability, resolution, environmental effects and the stability of the instrument are combined. The certificate states the expanded uncertainty with the coverage factor (usually k = 2) and the approximate level of confidence. The smallest uncertainty a laboratory can report is limited by its calibration and measurement capability (CMC) in the accredited scope; no better uncertainty may be declared.

  • The uncertainty of the reference microphone and the calibrator
  • Ambient pressure and temperature (affect the calibrator's output)
  • Ambient noise and the instrument's noise floor
  • The coupling effect at the microphone connection
  • Frequency-response uncertainty
  • Repeatability

Certificate Content

  • Laboratory details and certificate number
  • The customer and identification of the instrument calibrated: make, model, serial number
  • Date of calibration and date of issue of the certificate
  • The calibration method applied or reference to the procedure
  • Reference standards used and traceability information
  • Ambient conditions during calibration (temperature, humidity and, where relevant, pressure)
  • Measurement points, reference values and instrument readings
  • The deviations calculated
  • Expanded measurement uncertainty and coverage factor
  • The personnel who performed and who approved the calibration
  • The instrument class and whether it is retained
  • The frequency- and time-weighting results
  • The ambient pressure during calibration
  • The microphone and preamplifier serial numbers (the chain is calibrated as a whole)

Conformity Statement and Limits

A calibration certificate is NOT a certificate of conformity. Calibration measures and reports the deviation of the instrument; it does not correct that deviation and does not say that the instrument is "correct". To decide whether an instrument is fit for use, a tolerance (acceptance criterion) must be defined — and it is the user who sets that tolerance, not the laboratory. If a statement of conformity is requested, ISO/IEC 17025 clause 7.8.6 requires a DECISION RULE to be agreed in advance, defining how measurement uncertainty is taken into account in the decision; a conformity statement made without regard to uncertainty is misleading. The certificate also shows the condition AT THE TIME of calibration: later use, transport or shock can change the instrument's condition, and the certificate does not cover that. Where a next calibration date appears on the certificate, it is not the laboratory's decision but a reminder entered according to the interval stated by the customer.

Accreditation

Calibration is carried out under ISO/IEC 17025, which requires the laboratory's technical competence, impartiality and metrological traceability. Accreditation is granted per METHOD and per MEASUREMENT RANGE: a quantity or range that does not appear in the laboratory's scope annex is not accredited, and no certificate bearing the accreditation mark may be issued for that measurement. The scope also states the smallest uncertainty the laboratory can offer (CMC); no better uncertainty may be declared. EUROLAB holds internationally recognised accreditation under ISO/IEC 17025, granted by accreditation bodies that are signatories to the ILAC mutual recognition arrangement; whether your instrument can be calibrated within accredited scope is confirmed in writing at the quotation stage.

Scope and authorisations

Related Calibrations

Frequently Asked Questions

Is the check I do with a field calibrator enough?

No, it is an interim check. A single-frequency check with a calibrator before and after measurement shows the instrument is working at that moment, but does not verify frequency response, linearity or weighting behaviour. It does not substitute for periodic laboratory calibration — and the calibrator itself must also be calibrated.

Can I use the microphone with a different instrument?

No. In acoustic measurement, the microphone, preamplifier and main instrument are calibrated as a CHAIN; the certificate belongs to this combination. If a component is changed, the calibration ceases to be valid.

Why does ambient pressure affect the result?

The sound pressure level an acoustic calibrator produces depends on ambient pressure. At high altitude or a different air pressure, the same calibrator produces a different level; pressure is therefore measured and a correction applied during calibration.

Does calibration correct my instrument?

No. Calibration measures and reports the deviation; correcting it is a separate operation and must be requested separately.

Will the certificate state pass/fail?

Only if you stated a tolerance and decision rule. A calibration certificate reports deviation and uncertainty by default; for a statement of conformity you must give your acceptance criterion in advance and agree how measurement uncertainty is reflected in the decision.